Accuracy classes and maximum permissible errors for trade measuring instruments (EU MID)

Accuracy classes and maximum permissible errors that Directive 2014/32/EU (the Measuring Instruments Directive) assigns to instruments used for trade in the EU. Each record is one instrument type at one accuracy class under one measurement condition, with the permitted error the directive states. Covers the ten instrument-specific annexes: water meters (MI-001), gas meters and volume conversion devices (MI-002), active electrical energy meters (MI-003), thermal energy meters (MI-004), measuring systems for liquids other than water including fuel dispensers and road tankers (MI-005), automatic weighing instruments (MI-006), taximeters (MI-007), material measures (MI-008), dimensional measuring instruments (MI-009) and exhaust gas analysers (MI-010). Answers 'what is the maximum permissible error for a water meter', 'what accuracy class does a petrol pump have to be', 'MID accuracy class for a taximeter', 'allowed error on a weighbridge', 'thermal energy meter accuracy class 2 MPE' and 'is my fuel dispenser within tolerance'. The values are condition-dependent -- the error allowed at low flow differs from the error allowed at high flow, and in-service limits differ from new-instrument limits -- so a single percentage figure is almost always the wrong answer. A deliberate mirror: there is one original, but the numbers sit in ten annexes of a 300,000-character legal text with no per-instrument row structure, and search engines currently answer this question from document-sharing sites and equipment vendors rather than from the directive.

Records70
Sources1
Verified2026-08-05
Review by2027-08-05
LicenceEU legislation reused under Commission Decision 2011/833/EU on the reuse of documents from the EU institutions. legislation.gov.uk content is Open Government Licence v3.0. Note the single-database volume advisory: one publisher supplies every record here, which is the EU/UK database right question -- mitigated by the source being legislation published under an explicit reuse decision.

The data

Instrument typeAnnexAccuracy classConditionMaximum permissible errorScope or member state option
Gas meterIV1.0After durability testVariation not exceeding one-third of MPEMI-002. Class 1.0 durability test variation.
Water meterIIIN/A (single class)After durability test, lower zone (Q1 to Q2)Variation not exceeding 3 % of metered volumeMI-001. Durability test variation, lower zone.
Water meterIIIN/A (single class)After durability test, upper zone (Q2 to Q4)Variation not exceeding 1.5 % of metered volumeMI-001. Durability test variation, upper zone.
Gas meterIV1.5After durability test, upper zone (Qt to Qmax)Variation not exceeding 2 %MI-002. Class 1.5 durability test variation.
Liquid measuring system (aircraft refuelling)VII0.5Aircraft refuelling systems0.5 %MI-005. Accuracy class for aircraft refuelling.
Automatic catchweigherVIIIXI, XII, XIII, XIIII / Y(I), Y(II), Y(a), Y(b)All loads, automatic operationPer Tables 1-4MI-006. Eight accuracy classes for catchweighers.
Gas meters and volume conversion devices — volume conversion deviceAnnex IV (MI-002)ambient temperature 20 °C ± 3 °C, ambient humidity 60 % ± 15 %, nominal values for power supply± 0,5 %
Dimensional measuring instruments — area measuring instrumentsAnnex XI (MI-009)± 1,0 %, but not less than 1 dm²
Dimensional measuring instruments — multidimensional measuring instrumentsAnnex XI (MI-009)± 1,0 dd is the scale interval
Area measuring instrumentXIN/A (single class)Area of irregular shaped objects (e.g. leather)1.0 %, but not less than 1 dm2MI-009. Area measuring instrument MPE.
Material measures of lengthAnnex X (MI-008)Ibetween two non-consecutive scale marks± (0,1 + 0,1 L) mm, where L is length rounded up to next whole metre; c = 0,1 mm for terminal interval bounded by a surface
Material measure of lengthXI, II, III, D, SBetween two non-consecutive scale marksFormula (a + bL) mm; class-dependent a and b valuesMI-008. Length measure MPE formula.
TaximetersAnnex IX (MI-007)calculation of the fare± 0,1 %, minimum including rounding: corresponding to the least significant digit of the fare indicationexcluding any errors due to application of the taximeter in a taxi
Thermal energy meters — calculatorAnnex VI (MI-004)calculator sub-assemblyE c = ± (0,5 + Δθmin/Δθ)
Exhaust gas analysersAnnex XII (MI-010)0CO fraction± 0,03 % vol or ± 5 %, whichever is greater
Exhaust gas analyserXII0CO fraction+/- 0.03 % vol or +/- 5 %, whichever is greaterMI-010. Class 0 CO fraction MPE.
Exhaust gas analyserXII0 and ICO2 fraction+/- 0.5 % vol or +/- 5 %, whichever is greaterMI-010. CO2 fraction MPE.
Thermal energy meterVI1, 2, 3Complete instrument, rated operating conditionsClass-dependent formulaMI-004. Three accuracy classes.
Thermal energy meterVI1, 2, 3Complete instrumentFormula-based per class (Ef + Et + Ec)MI-004. MPE as percentage of true value.
Thermal energy metersAnnex VI (MI-004)1complete thermal energy meterE = ± (2 + 4 Δθmin/Δθ + 0,01 qp/q)E f , E t , E c according to points 7.1 to 7.3
Liquid measuring system (cryogenic)VII2.5Cryogenic liquids (temperature below -153 degrees C)2.5 %MI-005. Accuracy class for cryogenic systems.
Active electrical energy meterVA, B, CDifferent temperature rangesPer Table 2, relevant range appliesMI-003. Temperature range selection.
TaximetersAnnex IX (MI-007)distance travelled± 0,2 %, minimum value of mpe: 4 mexcluding any errors due to application of the taximeter in a taxi
TaximeterIXN/A (single class)Distance travelled measurement+/- 0.2 %, minimum value 4 mMI-007. Taximeter distance MPE.
Exhaust gas analyserXII0 and IEach fraction measured, rated operating conditionsGreater of absolute value or percentage of true value (per Table 2)MI-010. Greater-of rule for MPEs.
Multidimensional measuring instrumentXIN/A (single class)Edge length of smallest enclosing rectangular parallelepiped+/- 1.0 d (scale interval)MI-009. Multidimensional measuring instrument MPE.
Gas meterIV1.5Error after durability testNot exceeding twice the MPE in point 2MI-002. Class 1.5 error after durability.
Water meterIIIN/A (single class)Error after durability test, lower zone (Q1 to Q2)+/- 6 % of metered volumeMI-001. Error after durability, lower zone.
Water meterIIIN/A (single class)Error after durability test, upper zone (Q2 to Q4), cold water (0.1-30 degrees C)+/- 2.5 % of metered volumeMI-001. Error after durability, upper zone, cold water.
Water meterIIIN/A (single class)Error after durability test, upper zone (Q2 to Q4), hot water (30-90 degrees C)+/- 3.5 % of metered volumeMI-001. Error after durability, upper zone, hot water.
TaximeterIXN/A (single class)Fare calculation+/- 0.1 %, minimum corresponding to least significant digitMI-007. Taximeter fare calculation MPE.
Thermal energy meters — flow sensorAnnex VI (MI-004)1flow sensor sub-assemblyE f = ± (1 + 0,01 qp/q), but not more than 5 %
Fuel dispenser (not liquefied gases)VII0.5Fuel dispensers for non-liquefied-gas fuels0.5 %MI-005. Accuracy class for fuel dispensers.
Gas meterIVN/AGas meter with temperature conversion, 30 degrees C rangeMPE increased by 0.5 %MI-002. Temperature conversion additional MPE.
Exhaust gas analysersAnnex XII (MI-010)0HC fraction± 10 ppm vol or ± 5 %, whichever is greater
Exhaust gas analyserXII0HC fraction+/- 10 ppm vol or +/- 5 %, whichever is greaterMI-010. Class 0 HC fraction MPE.
Exhaust gas analyserXIIN/ALambda value calculation0.3 %MI-010. Lambda calculation MPE.
Liquid measuring system (liquefied carbon dioxide)VII1.5Liquefied carbon dioxide1.5 %MI-005. Accuracy class for liquefied CO2.
Liquid measuring system (liquefied gases, >= -10 C)VII1.0Liquefied gases under pressure, temperature >= -10 degrees C1.0 %MI-005. Accuracy class for liquefied gas systems.
Liquid measuring system (liquefied gases, < -10 C)VII1.5Liquefied gases under pressure, temperature below -10 degrees C (not cryogenic)1.5 %MI-005. Accuracy class for cold liquefied gas systems.
Water meterIIIN/A (single class)Lower zone (Q1 to Q2 excluded), any water temperature5 %MI-001. Volume measurement, lower zone.
Gas meterIV1.5Lower zone (Qmin to Qt), volume or mass at metering conditions3 %MI-002. Class 1.5 gas meter, lower zone MPE.
Length measuring instrument (for rope-type materials)XIIMeasurement of textiles, bands, cables, etc.0.125 %, but not less than 0.005 LmMI-009. Length measuring instrument Class I.
Liquid measuring system (milk)VII0.5Milk measuring systems0.5 %MI-005. Accuracy class for milk systems.
Exhaust gas analyserXII0 and IO2 fraction+/- 0.1 % vol or +/- 5 %, whichever is greaterMI-010. O2 fraction MPE.
Gas meters and volume conversion devices — volume conversion deviceAnnex IV (MI-002)other conversion devices at rated operating conditions± 1 %
Liquid measuring system (pipeline)VII0.3Pipeline measuring systems0.3 %MI-005. Accuracy class for pipeline systems.
Water metersAnnex III (MI-001)Q1 ≤ Q < Q2± 5 %
Water metersAnnex III (MI-001)Q2 ≤ Q ≤ Q4, water temperature > 30 °C± 3 %
Liquid measuring system (other than water)VII0.3, 0.5, 1.0, 1.5, 2.5Quantities >= 2 litres, measuring systems0.3 % to 2.5 % depending on accuracy classMI-005. Five accuracy classes for measuring systems and meters.
Active electrical energy meterVA, B, CRated operating conditions at varying load currentPer Table 2MI-003. MPEs per class, temperature range, and load.
Volume conversion device (other type)IVN/ARated operating conditions1 %MI-002. Other conversion device MPE.
Volume conversion device (temperature type)IVN/ARated operating conditions0.7 %MI-002. Temperature conversion device MPE.
Active electrical energy meterVA, B, CRated operating conditionsClass-dependent, per Table 2 of Annex VMI-003. Three class indices defined.
Exhaust gas analyserXII0, IRated operating conditionsPer Table 2 of Annex XIIMI-010. Two classes defined.
TaximeterIXN/A (single class)Real-time clock timekeeping0.02 %MI-007. Taximeter clock accuracy.
Volume conversion deviceIVN/AReference conditions (20 degrees C +/- 3, 60% humidity +/- 15%)0.5 %MI-002. Volume conversion device at reference conditions.
Liquid measuring system (road tankers, low viscosity)VII0.5Road tanker systems, liquids of low viscosity (< 20 mPa.s)0.5 %MI-005. Accuracy class for road tanker systems.
Automatic weighing instruments — automatic gravimetric filling instrumentsAnnex VIII (MI-006)Ref(x)static loads0,312 of the maximum permissible deviation of each fill from the average, as specified in Table 5, multiplied by the class designation factor (x)
Automatic gravimetric filling instrumentVIIIRef(x), X(x)Static loads, rated operating conditions0.312 of max permissible deviation times factor (x)MI-006. Filling instrument static weighing error.
Gas meters and volume conversion devicesAnnex IV (MI-002)1,5temperature conversion, 30 °C range around specified temperatureMPE increased by 0,5 %For a gas meter with temperature conversion, which only indicates the converted volume, the MPE of the meter is increased by 0,5 % in a range of 30 °C extending symmetrically around the temperature specified by the manufacturer that lies between 15 °C and 25 °C.
Gas meters and volume conversion devices — volume conversion deviceAnnex IV (MI-002)temperature conversion devices at rated operating conditions± 0,7 %
Thermal energy meters — temperature sensor pairAnnex VI (MI-004)temperature sensor pair sub-assemblyE t = ± (0,5 + 3 Δθmin/Δθ)
TaximetersAnnex IX (MI-007)time elapsed± 0,1 %, minimum value of mpe: 0,2 sexcluding any errors due to application of the taximeter in a taxi
TaximeterIXN/A (single class)Time elapsed measurement+/- 0.1 %, minimum value 0.2 sMI-007. Taximeter time MPE.
Continuous totaliserVIII0.5, 1, 2Totalised load+/- 0.25 % to +/- 1.0 % depending on classMI-006. Three accuracy classes for continuous totalisers.
Discontinuous totaliserVIII0.2, 0.5, 1, 2Totalised load+/- 0.10 % to +/- 1.00 % depending on classMI-006. Four accuracy classes for discontinuous totalisers.
Water meterIIIN/A (single class)Upper zone (Q2 included to Q4), water temperature <= 30 degrees C2 %MI-001. Volume measurement, upper zone, cold water.
Gas meterIV1.5Upper zone (Qt to Qmax), volume or mass at metering conditions1.5 %MI-002. Class 1.5 gas meter, upper zone MPE.
Automatic rail weighbridgeVIII0.2, 0.5, 1, 2Weighing-in-motion, single wagon or total train+/- 0.1 % to +/- 1.0 % depending on classMI-006. Four accuracy classes for rail weighbridges.

Where this came from

Every record above links the page it was taken from and quotes the sentence that states it. These are the 1 source this dataset was assembled from.

Machine-readable

70 records. last verified against source 2026-08-05. due for re-check by 2027-08-05.

Licence. EU legislation reused under Commission Decision 2011/833/EU on the reuse of documents from the EU institutions. legislation.gov.uk content is Open Government Licence v3.0. Note the single-database volume advisory: one publisher supplies every record here, which is the EU/UK database right question -- mitigated by the source being legislation published under an explicit reuse decision.