Gas meters and volume conversion devices · Annex IV · gas meter with temperature conversion, 30 °C range around the specified temperature
For Gas meters and volume conversion devices, annex is Annex IV; condition is gas meter with temperature conversion, 30 °C range around the specified temperature; maximum permissible error is increased by 0,5 %, recorded from its source on 2026-09-01.
- Instrument type
- Gas meters and volume conversion devices verified
- Annex
- Annex IV verified
- Condition
- gas meter with temperature conversion, 30 °C range around the specified temperature our reading
- Maximum permissible error
- increased by 0,5 % verified
Values marked our reading are our classification of what the source says — the source does not print them in those words. The quote below is the evidence for each one; judge it yourself.
What the source says
For a gas meter with temperature conversion, which only indicates the converted volume, the MPE of the meter is increased by 0,5 % in a range of 30 °C extending symmetrically around the temperature specified by the manufacturer that lies between 15 °C and 25 °C. Outside this range, an additional increase of 0,5 % is permitted in each interval of 10 °C.
— eur-lex.europa.eu, retrieved 2026-09-01
Where each value comes from
This source states these in separate places, so each value is shown with the passage that states it.
Instrument type
ANNEX IV GAS METERS AND VOLUME CONVERSION DEVICES (MI-002) The relevant requirements of Annex I, the specific requirements of this Annex and the conformity assessment procedures listed in this Annex, apply to gas meters and volume conversion devices defined below, intended for residential, commercial and light industrial use.
Annex
ANNEX IV GAS METERS AND VOLUME CONVERSION DEVICES (MI-002)
— all from eur-lex.europa.eu, retrieved 2026-09-01
Source
- eur-lex.europa.euhttps://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:32014L0032