Thermal energy meters · Annex VI · flow sensor sub-assembly, class 3
For Thermal energy meters, annex is Annex VI; condition is flow sensor sub-assembly, class 3; accuracy class is 3; maximum permissible error is not more than 5 %; scope or member state option is Where a Member State imposes measurement of residential use, it shall allow such measurement to be performed by means of any Class 3 meter, recorded from its source on 2026-09-01.
- Instrument type
- Thermal energy meters verified
- Annex
- Annex VI verified
- Accuracy class
- 3 verified
- Condition
- flow sensor sub-assembly, class 3 our reading
- Maximum permissible error
- not more than 5 % verified
- Scope or member state option
- Where a Member State imposes measurement of residential use, it shall allow such measurement to be performed by means of any Class 3 meter. verified
Values marked our reading are our classification of what the source says — the source does not print them in those words. The quote below is the evidence for each one; judge it yourself.
What the source says
Accuracy class
The source prints this value inside the passage below but does not name it there. The passage shows where it appears; what ties it to this record is the position on the page, not anything the passage says.
7.1. The relative MPE of the flow sensor, expressed in %, for accuracy classes: — | Class 1: , but not more than 5 %, | — | Class 2: , but not more than 5 %, | — | Class 3: , but not more than 5 %,
— eur-lex.europa.eu, retrieved 2026-09-01
Where each value comes from
This source states these in separate places, so each value is shown with the passage that states it.
Instrument type
ANNEX VI THERMAL ENERGY METERS (MI-004) The relevant requirements of Annex I, the specific requirements and the conformity assessment procedures listed in this Annex, apply to thermal energy meters defined below, intended for residential, commercial and light industrial use.
Annex
ANNEX VI THERMAL ENERGY METERS (MI-004)
Maximum permissible error
7.1. The relative MPE of the flow sensor, expressed in %, for accuracy classes: — | Class 1: , but not more than 5 %, | — | Class 2: , but not more than 5 %, | — | Class 3: , but not more than 5 %,
Scope or member state option
Where a Member State imposes measurement of residential use, it shall allow such measurement to be performed by means of any Class 3 meter.
— all from eur-lex.europa.eu, retrieved 2026-09-01
Source
- eur-lex.europa.euhttps://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:32014L0032