Reference Source

Thermal energy meters — temperature sensor pair

For Thermal energy meters, instrument type is Thermal energy meters — temperature sensor pair; annex is Annex VI (MI-004); condition is temperature sensor pair sub-assembly; maximum permissible error is E t = ± (0,5 + 3 Δθmin/Δθ), recorded from its source on 2026-08-05.

Instrument type
Thermal energy meters — temperature sensor pair
Annex
Annex VI (MI-004)
Condition
temperature sensor pair sub-assembly our reading
Maximum permissible error
E t = ± (0,5 + 3 Δθmin/Δθ)
Sourceeur-lex.europa.eu
Verified2026-08-05
Review by2027-08-05
DatasetAccuracy classes and maximum permissible errors for trade measuring instruments (EU MID)

Values marked our reading are our classification of what the source says — the source does not print them in those words. The quote below is the evidence for each one; judge it yourself.

What the source says

The relative MPE of the temperature sensor pair, expressed in %:

eur-lex.europa.eu, retrieved 2026-08-05

Source

Last verified against source: 2026-08-05. Due for re-check by 2027-08-05. This page as Markdown · OKF bundle · full dataset as JSON.