Gas meter

For Gas meter, annex is IV; accuracy class is N/A; condition is Gas meter with temperature conversion, 30 degrees C range; maximum permissible error is MPE increased by 0.5 %; scope or member state option is MI-002. Temperature conversion additional MPE, recorded from its source on 2026-08-05.

Instrument type
Gas meter verified
Annex
IV
Accuracy class
N/A
Condition
Gas meter with temperature conversion, 30 degrees C range our reading
Maximum permissible error
MPE increased by 0.5 %
Scope or member state option
MI-002. Temperature conversion additional MPE. our reading
Sourceeur-lex.europa.eu
Verified2026-08-05
Review by2027-08-05
DatasetAccuracy classes and maximum permissible errors for trade measuring instruments (EU MID)

Values marked our reading are our classification of what the source says — the source does not print them in those words. The quote below is the evidence for each one; judge it yourself.

What the source says

For a gas meter with temperature conversion, which only indicates the converted volume, the MPE of the meter is increased by 0,5 % in a range of 30 °C extending symmetrically around the temperature specified by the manufacturer that lies between 15 °C and 25 °C. Outside this range, an additional increase of 0,5 % is permitted in each interval of 10 °C.

eur-lex.europa.eu, retrieved 2026-08-05